VnV

Potential Effect(s) of FailureSeverity (S)Potential Cause(s)Occurrence (O)Current Design ControlsDetection (D)RPNRecommended Action
Reduced lumen diameter, restenosis9Insufficient radial strength4Radial compression testing, FEM analysis4144Optimize strut geometry and material properties
Vessel injury, malapposition8Uneven balloon expansion or poor design5Balloon expansion simulation, bench testing5200Improve expansion symmetry and balloon compatibility
Embolism, restenosis, device failure10Cyclic arterial loading3Fatigue simulation and durability testing4120Improve fatigue-resistant design and material selection
Thrombosis, restenosis9Poor surface finish or material incompatibility3Biocompatibility testing per ISO 10993381Improve coating and surface treatment
Procedure failure, vessel trauma7Excessive stiffness5Trackability and flexibility testing4140Reduce profile and improve flexibility
Improper deployment8High crimping stress4Crimping simulation and dimensional inspection4128Optimize crimping parameters and strut thickness
Acute thrombosis10Inadequate coating or rough surface2Hemocompatibility testing480Enhance anti-thrombotic coating
Vessel blockage or inadequate treatment9Improper sizing or poor anchoring3Deployment testing and imaging analysis5135Improve anchoring and sizing guidance
Restenosis or toxicity8Coating inconsistency4Drug release characterization4128Improve coating uniformity process
Structural degradation8Aggressive physiological environment2Corrosion testing348Use corrosion-resistant alloys or polymers
Process StepPotential Failure ModePotential Effect(s)Potential Cause(s)Prevention ControlsDetection ControlsSODRPNRecommended Actions
Material Receiving & InspectionIncoming material out of specificationCompromised stent mechanical properties, coating failure, patient injurySupplier variation, inadequate incoming inspectionApproved supplier list, CoC verificationIncoming dimensional and material inspection93381Strengthen supplier audits and incoming testing
Laser Cutting (Stent Patterning)Incorrect strut geometryPoor radial strength, restenosis, thrombosisLaser misalignment, programming errorValidated laser parameters, preventive maintenanceVision inspection, dimensional inspection844128Real-time monitoring and tighter process controls
ElectropolishingOver/under polishingReduced fatigue life, particulate generationIncorrect voltage/time, electrolyte variationProcess parameter validationSurface finish and dimensional checks744112Automate parameter control and trend monitoring
CleaningResidual particulates or contaminantsCoating defects, thrombogenic riskInsufficient cleaning cycle, bath contaminationValidated cleaning procedureParticulate and TOC monitoring83496Increase particulate testing frequency
Coating Application (Drug & Polymer)Coating thickness out of specificationUnder/over dosing, coating delaminationViscosity variation, dip speed variationValidated coating processCoating thickness measurement844128Inline viscosity monitoring and automation
Drying & CuringIncomplete drying/curingPoor adhesion, particulate formationIncorrect temperature/timeOven cycle validationTemperature monitoring and adhesion testing73484Improve oven alarms and adhesion testing
Crimping & MountingStent deformation/fractureAcute closure, compromised deploymentExcessive crimping force, mandrel damageValidated crimping processPost-crimp visual inspection93381Vision-based strut integrity inspection
SterilizationSterilization dose out of specificationInfection risk, material degradationIncorrect cycle parametersValidated sterilization cycleDosimetry and release testing92354Routine calibration and preventive maintenance
Final Inspection & PackagingSeal failure or labeling errorLoss of sterility, wrong product useHuman error, sealing parameter variationValidated sealing processSeal integrity and label verification83372Increase label verification and seal testing